Sample Take-Off Arrangements

3 important questions on Sample Take-Off Arrangements

Where should a sample probe be located for a liquid samples?

Side of process line with arrangements for unplugging and flushing sample lines.

Where in a pipe should a sample probe be located?

Within the middle third of a pipe to avoid slower slipstream near pipe wall and any pipe wall effects like rust or memory effect.

Who decides sample take off location?

Joint decision between analyser and process technologists, to ensure a reliable fit for purpose design.